VTC-100PA is CE certified tabletop spin coating device dedicated to spin coat small substrates up to 100 mm diameter at Max. 10000 rpm. The spin head actuator is a precision DC servo motor with accurate speed and acceleration control. The substrate can be held by vacuum (for thinner objects). The device has a user-friendly keyboard and LCD display for programming various parameters up to 12 programs, 6 segments. Spin duration, spin speed, and acceleration can be programmed from the front panel. All components in the device are corrosion protected to make it cleanroom compatible.
Specifications | |
---|---|
Main Features | • 500 ~ 10000 RPM spinning speed programmable • Accept substrate from 5 mm to 5″ • PP chamber for anti-corrosive operation. • One gas inlet port on top of cover which can fill inert gas during coating • One 20ml Plastic Syringe is included • Suitable using Ar gas ( vacuum pump must be outside glovebox) • Safety interlock is integrated into the lid |
Vacuum Chuck & Vacuum pump | • Chamber size: 150 mm diameter • 2 sets of PP anti-corrosive vacuum chucks with different sizes (Φ44mm, Φ9mm), which can hold substrate from 5 mm to 5″ Dia. ( see pic below ) • Click the picture below right to watch demo video for how to use 9mm vacuum chuck hold 5mm sample. • One 120L/min Oilless Vacuum Pump is included in the standard package, please click the picture below for specification details. |
Programmable Control | • Duration up to 60 seconds • Acceleration: 100 – 2000 rpm / sec adjustable • Accuracy: < 5% • Digital control panel can store 12 programs and each program has 6 segments programmable for speeds, acceleration, and spin duration (please click picture below-right to see typical program setting) |
Input Power | • AC 220V (You may order a transformer to operate at 110V) • 1000 W total for oilless pump and spin cater |
Gas Inlet | 6mm diameter gas inlet port allows customers to fill desired gas into the chamber, which can make coating under variable gas flow environment. |
Warranty & Compliance | • One year limited with lifetime support • Rusting and damage due to improper storage condition or maintenance are not covered by warranty • CE certified |
Warnings | • Do not overflush the water bowl/base with any kinds of liquid. Excess liquid will leak into the motor shaft and cause damage. Please use fabric cloth to wipe out the residues remain in the bow/base. • If necessary, turn on the water outlet when flushing the water bowl/base to avoid liquid overflowing during the operation. • For more details about the cleaning procedure, please review the cleaning instruction for the spin coaters. |
Shipping Dimensions | 40″(L) x 30″(W) x 30″(H) |